Research on Reliability Evaluation of Plateau Emergency Rescue Ventilator based on Four Comprehensive
DOI:
https://doi.org/10.54691/rwxwa270Keywords:
Plateau Area; Four Comprehensive Environmental Tests; Ventilator; Environmental Adaptability.Abstract
The harsh natural environment of high-altitude regions leads to frequent emergencies. As a critical piece of equipment in emergency rescue, the reliability of ventilators directly impacts treatment efficiency and patient safety. This study aims to evaluate the ventilator's start-up capability, operational stability, and performance maintenance under typical high-altitude transportation conditions. Using the Chengdu-Lhasa high-altitude transportation route as the research background, a specific model of ventilator was selected as the research object. A comprehensive environmental testing platform was used to simulate the combined effects of four environmental stresses: temperature, humidity, vibration, and low air pressure. A full-process test operation covering pretreatment, initial testing, operational loading, performance monitoring, and recovery was conducted on the ventilator. The study found that the ventilator faces a risk of start-up failure under the coupled conditions of low air pressure and low temperature. However, during operation, its performance indicators (such as control accuracy) still meet requirements under vibration and low-pressure interference. It is recommended to optimize the ventilator's structural design and high-altitude environmental adaptability testing procedures to provide a basis for the selection, deployment, and standardization of high-altitude emergency medical equipment.
Downloads
References
[1] Gao Na, Yuan Lijuan. Comparative Analysis of Earthquake Rescue Situation in Tingri, Tibet and Yushu, Qinghai [J]. China Emergency Rescue, 2025, (03): 28-32.
[2] Hu Yifan, Tian Bingwei, Chang Ying. Characteristics, Difficulties and Coping Strategies of Rescue Operations in High-Altitude and Cold Environments-Taking the 6.8 Magnitude Earthquake in Tingri, Tibet as an Example [J]. China Emergency Rescue, 2025, (03): 33-39.
[3] Yao Qinqi, Tang Mingkang, Ye Longying, et al. Study on Evaluation and Screening Indicators of Emergency Treatment Ventilators [J]. Medical and Health Equipment, 2024, 45(07): 8-16.
[4] Hang Jianjin, Zhang Chaoqun, Wu Xiangyang, et al. A Brief Discussion on the Clinical Application and Quality Control Management of Ventilators in High-Altitude Environments [J]. Medical and Health Equipment, 2013, 34(9): 2.
[5] Hang Jianjin, Li Beiqiang, Zhang Chaoqun, et al. Practice of Ventilator Quality Testing in High-Altitude Environments [J]. Medical and Health Equipment, 2014, 35(10):2.
[6] Zhu Zifu, Ma Shengcai, Niu Hangduo, et al. Research progress of ventilators[J]. Chinese Journal of Medical Devices, 2021, 45(4):5.
[7] Wang Peng, Li Shu, Niu Dainan, et al. Design and test analysis of general vibration fixture based on vibration table[J]. Environmental Technology, 2025, 43(1):145-149.
[8] Feng Haibo, Xin Ying. Suppression and application of fixture modes in vibration test[J]. Mechanical Research and Application, 2023, 36(5):64-66.
[9] Shi Cheng, Zhang Liwei, Tan Haiqing, et al. Research on three-dimensional environmental test method of temperature, humidity and vibration[J]. Environmental Technology, 2024, 42(12):68-72.
[10] Lu Cailing, Kang Ningmin, Wang Shurong. Discussion on vibration test method of equipment transportation[J]. Environmental Technology, 2011, 33(02):13-17+48.
[11] Xu Yusheng. The Influence of Simulated Lungs with Different Compliance and Resistance on the Accuracy of Tidal Volume Output of Ventilators [J]. China Medical Device Information, 2023, 29(9):42-44.
[12] Wang Tianying, Gu Wei, Zhou Yang, et al. Comparative Study of Different Types of Oxygen Sensors for Ventilators [J]. China Medical Equipment, 2023, 20(4):177-179.
[13] Deng Ju. Discussion on the Calibration Method of Continuous Positive Pressure for Non-invasive Ventilators [J]. Metrology and Testing Technology, 2024, 50(11):118-120.
[14] Zhang Xuexia. The Effect of Non-invasive Ventilator-Assisted Treatment on Neonatal Respiratory Distress Syndrome and its Influence on Blood Gas Indicators and Ventilator Parameters of Infants [J]. China Medical Innovation, 2024, 21(18):67-71.
[15] Wu Yunzhen, Liu Xinlei. Exploring a method for estimating dynamic intrinsic positive end-expiratory pressure during invasive mechanical ventilation [J]. Chinese Journal of Critical Care Medicine, 2024, 36(01):86-89.
Downloads
Published
Issue
Section
License
Copyright (c) 2025 Scientific Journal of Intelligent Systems Research

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.